The T274 is a new microprocessor-based portable calibrator from Gometrics Instruments, designed for calibrating process instruments. It measures and generates current (mA) and voltage (V/mV) signals, combining the functions of the T224 current calibrator and the T234 voltage calibrator—which it replaces—into a single instrument.
It is part of the Tinycal line, which has been updated in both design and performance: resolution and accuracy have been improved compared to previous models, with enhancements such as a backlit display, better long-term stability, and the ability to add accessories, such as multi-input connectors or probes.
On the front panel, the measurement is selected using the IN/OUT key; the display shows the value to be measured at all times, and the generation of setpoints can be configured in engineering units or as a percentage, with programmable intervals ranging from 1 to 30 seconds that can be stored in its 6 preset keys (S1–S6).
The T274 features a built-in 24 Vdc power supply that allows the loop to be powered directly from the calibrator (active mode) or to operate in passive mode. With an accuracy of 0.015% FS and a one-year uncertainty of 0.025% FS, it offers high-precision, high-resolution measurements with the long-term stability characteristic of the Tinycal series.
Its high reliability, ease of use and programming, as well as its light weight and compact size, make it a very useful standard tool in an instrumentation workshop, whether in the field or on-site.
| Accuracy | 0.015% FS |
| Uncertainty | 0.025% FS |
| Loop Power Supply | 24 Vdc, maximum 30 mA |
| External power supply | maximum 35 Vdc |
| Overvoltage protection | maximum 60 V, 500 mA |
| Preliminary Selection Reports | 6 (S1…S6) |
| Keyboard | 14 membrane keys |
| Storage Battery | Rechargeable NiMH battery pack, 3.6 V, 2000 mA/h |
| Battery life — using internal power supply at 12 mA | Typical: 16 hours |
| Battery life — without using the internal power source | Typical: 50 hours |
| Output range | 0 to 30 mA |
| Resolution (0 to 20 mA) | 0.001 mA / 0.01% |
| Resolution (20 to 30 mA) | 0.01 mA / 0.01% |
| Maximum loop impedance | 1100 Ω at 20 mA |
| Generation range | 0 to 20 V / 0 to 200 mV |
| Resolution | 0.001 V / 0.01 mV / 0.01% |
* Includes nonlinearity, hysteresis, repeatability, thermal drift between 20 °C and 26 °C, and stability over 1 year.
| Measurement range | 0 to 30 mA |
| Resolution (0 to 20 mA) | 0.001 mA / 0.01% |
| Resolution (20 to 30 mA) | 0.01 mA / 0.01% |
| Input impedance | 10 Ω |
| Generation range | 0 to 30 V / 0 to 200 mV |
| Resolution (0 to 20 V) | 0.001 V / 0.01% |
| Resolution (20 to 30 V) | 0.01 V / 0.01% |
| Resolution (or at 200 mV) | 0.01 mV / 0.01% |
* Includes nonlinearity, hysteresis, repeatability, thermal drift between 20 °C and 26 °C, and stability over 1 year.