The TINYCAL series instruments offer a dual function: generation/simulation and measurement. The instrument’s keypad allows its configuration for the generation of fixed values, both in engineering and percentage values. The keypad also allows the generation by jumps as well as the configuration of the jump time between 1 and 9 seconds for each section, for which the instrument is provided with 6 keys to store the configuration values. The simulation from the instrument – T240 and T250 – offers the possibility to configure it to simulate the response of the sensor (Pt100 or type of thermocouple selected).
The measurement function, which is easily selectable by means of the IN/OUT key and whose characteristic is clearly indicated on the instrument display, allows the indication of the value to be measured. The measuring circuit is protected against short-circuits and external voltages up to 30 V, thus ensuring a longer service life of the instrument. The T220 and T225 models of the TINYCAL family are equipped with an integral 24 Vdc power supply, which allows, by means of the appropriate connection, the loop to be powered from the calibrator itself (active mode) or voluntarily to work in passive mode.
All the instruments of the TINYCAL family offer great precision and high resolution, which together with their ease of use, great reliability and ease of transport, due to their reduced weight and dimensions, make these instruments basic standards in the instrumentation workshop.
Technical data |
| Common technical features in the Tinycal Series | |
| Operating temperature | 0 to 50 ºC |
| Temperature coefficient | 0.003% /ºC |
| Loop power supply | 24 Vdc (models T220 / T225) |
| Rechargeable batteries | NiMH |
| Autonomy | 16 hours |
| Power supply to battery charger | 230 V 50 Hz |
| Display | LCD, 41/2 digits 10 mm high |
| Dimensions | 83 x 152 x 33 mm |
| Weight | 240 to 300 gr. (depending on model) |
| Generation | Manual or Automatic |
| Jump time | Programmable between 1 and 9 seconds |
| Marking | CE |
| Warranty | 2 years |
| Signal | 4-20 mA |
| Range | 0-24 mA |
| Input impedance | 10 Ω |
| Accuracy at 23ºC ± 3ºC | ± 0.015% FS |
| Uncertainty at 1 year | ± 0.025% FS |
| Resolution | 0.001/0.01 mA |
| Function | Measurement and generation in active or passive form selectable by connection |
| Autonomy | 150 h in measurement; 11 h generating 20 mA with internal source |
| Signal | 10 – 50 mA |
| Range | 0 – 51 mA |
| Signal | 0-10 V dc |
| Range | 0-12,000 V |
| Input impedance | >1 M Ω |
| Accuracy at 23ºC ± 3ºC | ± 0.015% FS |
| Uncertainty at 1 year | ± 0.025% FS |
| Resolution | 0,001 V |
| Function | Measurement and generation |
| Signal | Resistance and Pt100 (according to IEC 751) |
| Simulation range | 80 to 390 Ohms (-50 to 850 ºC) (Current in simulation from 0.1 to 2 mA) |
| Simulation response | 2 mS (accepts pulsed transmitters/indicators) |
| Accuracy at 23ºC ± 3ºC | -50 to 150 ºC: 0.1 ºC (up to 850 ºC: 0.2 ºC) |
| Uncertainty at 1 year | -50 to 150 ºC: 0.2 ºC (up to 850 ºC: 0.4 ºC) |
| Measuring range | 20 to 320 Ohms (-220 to 600 ºC) (applied current 1mA approx.) |
| Input impedance | >10 M Ω |
| Accuracy at 23ºC ± 3ºC | -220 to 150 ºC: 0.1 ºC (± 1 count); up to 600 ºC: 0.2 ºC (± 1 count) |
| Uncertainty at 1 year | 220 to 150 ºC: 0.2 ºC (± 1 count); up to 600 ºC: 0.4 ºC (± 1 count) |
| Resolution | 0,1 Ohm or 0,1 ºC/0,2 ºC |
| Function | Measurement and Generation/Simulation |
| Signal | Thermocouple |
| Type | E, J, K, R, S, T (acc. to IEC 584-1) |
| Range | -15.00 to 75.00 mV and equivalence in ºC according to thermocouple:
|
| Input impedance | > 10 MΩ |
| Accuracy at 23ºC ±3ºC | ± 0.025% FS |
| Uncertainty at 1 year | ± 0.05% FS |
| Resolution | 0.01 mV and 1ºC |
| Function | Measurement and Generation/ Simulation |
| Cold junction compensation | Internal or manual from 0 to 50ºC |
| Signal | pH |
| Ranges | -450 to +450 mV; 0.00 to 14.00 pH |
| Input impedance | > 1000 GΩ |
| Accuracy at 23ºC ±3ºC | ± 0.025% FS |
| Uncertainty at 1 year | ± 0.05% FS |
| Includes | CHECK function to verify the proper input impedance of the PH meter equipment. |
| Resolution | 0.01 mV and 1ºC |
| Function | Measurement and generation/ Simulation |