ADVANTAGES
|
APPLICATIONS
|
CHARACTERISTICS
|
Technical specifications |
Pressure range
- ± 50 Pa (Model 1)
- ± 150 Pa (Model 2)
- ± 500 Pa (Model 3)
- ± 2500 Pa (Model 4)
- ± 10000 Pa (Model 5)
- ± 20000 Pa (Model 6)
Optional outputs2 wires 4-20 mA3 wires: 0-1 Vcc to 0-10 Vcc4 wires: 0-1 Vcc to 0-10 Vdc, ± 1 Vdc to ± 10 Vdc and any of the mA and V options above for 4 isolated wiresDisplay (optional)Most common: differential pressure, volume flow, mass flow rate and speed unitsPerformance adjustment0.0 to 60.0 secondsSquare root functionStandardRelay levelOptional: 2 relays, 2A to 55 Vac, 30 VccZero controlOptional: automatic or remotePneumatic port1/4 “BSP HSPECIFICATIONSImproved accuracy at 20 ºC
- 10% to 100% range: <± (0.25% reading + 1 digit)
- 0 to 10% range: <± (0.025% range + 1 digit)
Standard accuracy at 20 ºC
- 10% to 100% range: <± (0.5% reading + 1 digit)
- 0 to 10% range: <± (0.05% range + 1 digit)
Span adjustment10% to 100% of the range (can be set in any range of the instrument, below 20% accuracy is reduced to standard specifications)Displacement0.2% per yearTemperature coefficientsStandard :
- Zero: <0.2% / ºC
- Range: <0.4% / ºC
Improved :
- Zero: <0.02% / ºC
- Range: <0.02% / ºC
Work temperature– 10 to 60 ºCOutput resolutionBetter than 0.033% SPANOverpressure20 x DP of the rangeStatic pressure-1 a +10 bar relativaMinimum response step100 msRefreshment of exit50 msBUILDINGSecurity boxIP66 PlicarbonateDimensions144 x 155 x 93 mmMaterial in contactAluminum, nikel, mica and PTFECompatibilityAir and other non-corrosive gases, maximum 95% non-condensed moistureWeight1.4 kg
CHARACTERISTICS |